Output Amplifier

The output amplifier low-pass filters the signal emerging from the Output Mixer and then amplifies it into a usable range. The amplifier also contains a detector for leveling the output.

Bias Board

The bias board contains the ALC circuitry for the tracking generator and distributes dc power from the Al4 Frequency Control Assembly to the rest of the tracking generator. The ALC inputs come from the Al4 Frequency Control Assembly (for controlling the power level), the EXT ALC INPUT on the rear panel, and the detector in the Output Amplifier. The ALC loop drives the Modulator.

The following troubleshooting information is aimed at isolating tracking-generator-related faults to either the A10 Tracking Generator Assembly, or one of the other supporting assemblies, such as A14, A15, or A7. The A10 Tracking Generator is not field-repairable; a rebuilt-exchange assembly is available.

Output Goes Unleveled (ERR 900 or ERR 901)

The ADC on the A3 Interface Assembly is used to monitor the control line ALC MON (ALC Monitor) from AlO. If ALC MON is greater than +1.091 Vdc or less than -0.545 Vdc, ERR 900 TG UNLVL will be displayed, indicating that the output of the tracking generator (or “TG”) is unleveled. The TG can typically be set for +2.8 dBm output power and remain leveled. In any case, the output should remain leveled for output power settings of less than the maximum leveled output power. Refer to Table l-l in the Installation and Verification Manual for more information on this specification.

It is normal for the TG to be unleveled at frequencies below 300 kHz. If the TG output is unleveled and the start frequency of the TG is below 300 kHz, ERR 901 TGFrqLmt (TG Frequency Limit) and ERR 900 TG UNLVL may be displayed (refer to General Troubleshooting for information on checking for multiple error messages). If the start frequency is changed to be greater than 300 kHz and the output is still unleveled, ERR 900 TG UNLVL will be displayed (see above).

The ALC MON line is monitored only at the end of a sweep. For this reason, it is possible that the output could be unleveled during a portion of a sweep, and, if the output returns to a leveled condition by the end of the sweep, ERR 900 TG UNLVL will not be displayed.

If ERR 900 TG UNLVL is displayed, proceed as follows:

1.Check at which frequencies the output is unleveled. Set the analyzer to zero span and step the center frequency in 50 MHz increments. Note at which frequencies the output is unleveled.

2.Check at which power levels the output is unleveled. Connect the RF OUT 5Ofl to the INPUT 5052. With the analyzer in zero span, set the Center Frequency to 300 MHz or one of the frequencies noted in step 1, with the analyzer in zero span. Press [AUXCTRL),

TRACKING GENRATOR, SRC PWR ON, MORE I OF 3, TRACKING PEAK. Wait for the

“PEAKING” message to disappear. Step the TRK GEN RF POWER setting in 1 dB increments and note at which power levels the output is unleveled. It is acceptable for the output to be unleveled only at power levels greater than the specified maximum leveled output power level.