22.Change the reference level from
Log Offset/Log Expand
See function block X of A3 Interface Assembly Schematic Diagram (sheet 5 of 6).
The log scales are modified using a combination of amplification and digital trace manipulation. The video input to the A3 assembly is either 10 dB/div or linear. To obtain the 5 dB/div scale, the CPU manipulates the trace data from the 10 dB/div scale. To obtain the 2 dB/div scale, the video signal is amplified and offset so that
In 2 dB/div, Log Offset/Log Expand amplifies the top 20 dB of the display. This is done by offsetting the video signal by
1. On the HP 8560A press (PRESET),
2.Disconnect W26 (coax 2) from A3JlOl and connect the output of a function generator to A3JlOl.
3.Set the function generator to the following settings:
Output . . . . . . . . . | ..~.......................................... | Sine wave |
Amplitude . . . . . . . . | . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . | 1V |
DC Offset . . . . . . | . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . | . . t500 mV |
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . | . . . . . 50 Hz |
4.Set the HP 8560A sweep time to 50 ms.
5.Adjust the function generator amplitude and offset until the sine wave fills the entire graticule area.
6.Measure and note the function generator’s
v(10 dB/div) = |
| V |
|
7.Set the HP 8560A’s dB/div to 2 dB.
8.Readjust the function generator amplitude and offset until the sine wave again fills the entire graticule area.
9.Measure the function generator’s
v(2 dB/div) = |
| V |
|
10.The ratio of voltage recorded in step 6 to the voltage recorded in step 9 should be 5 f3%. If the ratio is not 5, troubleshoot the A3 Interface assembly.
11.Reconnect W26 to A3JlOl.