A10 Tracking Generator (Option 002)

A10 Tracking Generator (Option 002)

Removal

1.Use a 5/16 inch wrench to remove the Tracking Generator’s RF OUT, LO OUT, and LO IN semi-rigid cables.

2.Disconnect W14 and W16 from the Tracking Generator.

3.Remove the three screws (1) securing the Tracking Generator to the center deck. These screws are located on the top side of the center deck as illustrated in Figure 3-17.

4.Remove the Tracking Generator and disconnect W48, coax 80.

CL

Q

( 3 P L A C E S )

Figure 3-17. A10 Tracking Generator Mounting Screws

Replacement

1.Connect W48, coax 80, to the Tracking Generator’s INPUT connector.

2.Orient the Tracking Generator so that its LO IN, LO OUT, and RF OUT connectors are closest to the Al3 Second Converter.

3.Loosely connect the LO IN, LO OUT, and RF OUT semi-rigid cables.

4.Secure the Tracking Generator to the analyzer’s center deck using the three screws removed in step 3 of the removal procedure.

5.Tighten the semi-rigid cables.

6.Connect W14 and W16 to the Tracking Generator.

Assembly Replacement 3-31