All YTO Spanwidth | Sweep Applied To |
20.1 MHz to 3.8107 GHz | All YTO’s main coil |
1.01 MHz to 20.0 MHz | All YTO’s FM coil |
100 kHz to 1 MHz | Roller Oscillator PLL’s Main Oscillator |
100 Hz to 100 kHz | Roller Oscillator PLL’s Offset Oscillator |
When the sweep ramp is applied to one of the PLLs, the analyzer must prevent this loop from trying to compensate for changes in the output frequency. To accomplish this, the analyzer breaks the PLL by switching the output of the PLL’s phase detector to ground.
Reference PLL (P/O A15)
The 600 MHz Reference PLL provides 600 MHz for the second LO, 300 MHz for the third LO, and the Sampling Oscillator reference. The PLL is locked to a 10 MHz TCXO
YTO PLL (A7, All, P/O A14, P/O A15)
The YTO PLL produces the instrument’s first LO (3.0 to 6.81 GHz). The oscillator’s output is sampled by the Sampling Oscillator, and the resulting frequency is
The A15A2 Sampler mixes the LO signal from the A7 LODA with a harmonic of the Sampling Oscillator. The mixing product, the Sampler IF, is between 63 and 105.5 MHz (same frequency range as the Roller Oscillator PLL).
Offset Lock Loop (P/O A15)
The 280 MHz to 298 MHz Sampling Oscillator is used to sample the YTO. By changing the Offset Lock Loopls programmable dividers, the YTO frequency can be changed.