A2
CONTROLLER
(TAM TEST
SPi7
CONNECTOR)
Figure
10.Press SOFT KEY #4. The results will be sent to the printer.
11.If a failure is indicated in any of these tests, the fault lies on the A2 Controller assembly. To obtain more information:
a.Press the down arrow key one less time than the test number. (For example, press it twice for the third test on the list.)
b. Press SOFT KEY t3 , then SOFT KEY 64, and when the printout is complete, SOFT KEY #6.
12.If no failures were indicated in testing the A2 Controller, move the probe cable to A17J4.
13.Press SOFT KEY #I and wait five seconds.
14.Press SOFT KEY #4 . The results will be sent to the printer.
15.If no failure is indicated in the printout, refer to “High Voltage Supplies” in Chapter 12.
Controller Section