Removal
1.Remove the two screws securing the assembly to the analyzer’s center deck.
2.Use a
3.Disconnect W42 (W43 on Option 002) at the
4.Remove the gate bias wire, color code 97, and W12 from the A7 assembly.
5.Remove the assembly and disconnect W34.
Replacement
1.Use a
2.Connect gate bias wire, color code 97, to the A7 gate bias connection next to A7J2.
3.Connect cable W12 to the A7 assembly.
4.Place gate bias wire, color code 97, beneath W38 and connect W38 to A7Jl. Connect W42 (W43 on Option 002) to the
5.Use two panhead screws to secure A7 to the center deck. Be sure to attach the ground lug on the screw next to A7J4.
6.Tighten all RF cable connections.