11. Calibrator Amplitude Adjustment

3.Connect the HP 8482A through an adapter directly to the HP 8560A CAL OUTPUT connector.

4.Adjust A15R561 CAL AMPTD for a -10.00 dBm reading on the HP 8902A display.

Adjustment Procedures 2-49