6. Sampling Oscillator Adjustment
12.Record the displayed voltage in Table
13.Press (m) on the HP 8560A. Use the step keys to set the analyzer center frequency to the frequencies listed in Table
displayed voltage in the table. | ’ |
14.If the difference between the maximum and minimum voltages is less than 0.50 V, and all voltage readings are between to.5 and t2.5 Vdc, proceed to step 19.
15.Locate the center frequency at which the voltage is lowest. Use the II] and II] keys to set the HP 8560A to this frequency.
16.Readjust SMPL MATCH 1 to set the displayed voltage to 0.8 $0.1 Vdc.
17.Move the positive DVM test lead to A15J400 pin 3.
18.Set the HP 8560A center frequency to 511 MHz.
19.Readjust A15R237 SMPL PWR ADJ if necessary, until the voltage at A15J400 pin 3 is
20.Disconnect the DVM probes from A15J400.
| |
|
|
Note | This adjustment should be necessary only if the coaxial resonator 2200 has |
| been replaced or if there was insufficient range in the Sampling Oscillator |
| |
|
|
| FREQUENCY |
| COUNTER |
POWER |
|
SUPPLY | SPECTRUM |
| |
A 1 4 | ANALYZER |
S K 1 8
Figure
21.Set the HP 8560A m switch off and remove the top shield over the sampling oscillator. Connect the equipment as shown in Figure
22.Remove any existing shorts from the exposed
Adjustment Procedures