41.The Main Coil Coarse and Fine DACs correct any initial pretune errors in the YTO main coil. The DACs adjust the
42.Set the HP 8560A to the settings listed below. These set both DACs to 128 (the DAC setting range is 0 to 255).
C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...300 MHz SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OHz TRIG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..CONT
43. Verify the voltages listed in Table
Table
Measurement Points | Voltages | |
A14J17 pin 2 | Vdc | |
A14J17 pin 3 | Vdc | |
A14J17 pin 5 | +8.1 Vdc | |
|
|
|
44. Place jumper A14J23 in the NORMAL position.
Note | The All YTO has an initial pretune accuracy of f20 MHz. However, when |
| the Roller Oscillator frequency is <73 MHz this is changed to |
| If the Roller Oscillator frequency is >93 MHz, the accuracy is changed to |
| |
| the Sampler IF within the acquisition range of the YTO Loop. When dealing |
| with the Sampler IF and an unlocked YTO, the same frequency differences |
| apply to the Sampler IF. |
|
|
45. Set the HP 8560A to the following settings:
C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...300MHz SPAN . . . . . . . . . . . . . . . . . . . . . :. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Hz
46.Place jumper A14J23 in the TEST position.
47.Measure the output of the Main Coil Tune DAC (A14J18 pin 3) with a DVM. Refer to function block C of Al4 Frequency Control Schematic (sheet 2 of 5) in the
48.If the HP 8560A’s center frequency is 300 MHz, the voltage at A14J18 pin 3 should measure
V=
49.The voltage at A14U330 pin 2 should measure 3.4 V f0.2 Vdc. This represents a current setting the YTO to approximately 2.95 GHz.
Synthesizer Section