Table
Reference |
| Adjustment | Adjustment | Description |
Designator |
| Name | Number |
|
A15R561 C | A L A M P T D | 11 | Adjusts amplitude of the 300 MHz calibrator signal | |
|
|
|
| to |
A15R926 E | X T B I A S Z E R O | 14 | Adjusts zero bias point of external mixer bias. | |
A17R4 | Z GAIN | 2 | Adjusts maximum intensity. | |
A17Rll | CUTOFF | 2 | Adjusts intensity to turn off blanked lines. | |
A17R21 | Z | FOCUS | 2 | Adjusts focus for lines of different brightness. |
A17R26 | X | FOCUS | 2 | Adjusts focus at the left and right corners of the display |
A17R34 | COARSE FOCUS | 2 | Adjusts focus at the center of the display. | |
A17R55 | X | GAIN | 2 | Adjusts the |
A17R57 | X POSN | 2 | Adjusts the CRT horizontal position. | |
Al7R75 | Y | GAIN | 2 | Adjusts the |
A17R77 | Y POSN | 2 | Adjusts the CRT vertical position. | |
A17R90 | TRACE ALIGN | 2 | Adjusts the display axis rotation. | |
A17R92 | DDD | 2 | Adjusts focus of the center of the display. | |
417R93 | ASTIG | 2 | Adjusts for the spot roundness on the CRT display. | |
|
|
|
|
|
|
| Table |
|
|
|
Reference | Adjustment | Basis of Selection |
Designator | Number |
|
|
|
|
A5C204 | 3 | Selected to optimize center frequency of LC tank that loads the crystal. |
A5C216 | 3 | Selected to optimize center frequency of LC tank that loads the crystal. |
A5C326 | 3 | Selected to optimize LC pole center frequency. |
A5C327 | 3 | Selected to, optimize LC pole center frequency. |
A5C505 | 3 | Selected to optimize center frequency of LC tank that loads the crystal. |
A5C516 | 3 | Selected to optimize center frequency of LC tank that loads the crystal. |
A5C717 | 3 | Selected to optimize LC pole center frequency. |
A5C718 | 3 | Selected to optimize LC pole center frequency. |
A15U802 | 16 | Selected to set the gain of the second IF to 12 dB. |
|
|
|