7. YTO Adjustment
7. YTO AdjustmentAssembly Adjusted
Al4 Frequency Control Assembly
Related Performance Tests
Frequency Span Accuracy
Frequency Readout Accuracy and Frequency Count Marker Accuracy
Description
The YTO main coil adjustments are made with the phase-lock loops disabled. The YTO endpoints are adjusted to bring these points within the capture range of the main loop. The YTO FM coil is adjusted to place the 300 MHz CAL OUTPUT signal at the center vertical graticule in a 20 MHz span.
FREQUENCY
COUNTERSPECTRUM ANALYZER
A l 4 F R E Q C O N T R O
S K 1 9
A l 5 R F
Figure 2-10. YTO Adjustment Setup
Equipment
Microwave Frequency Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 5343A Option 001
Adapters
Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1250-1476 Type N (f) to APC 3.5 (f) (Option 026 only). . . . . . . . . . . . . . . . . . . . . . . . . .1250-1745 APC 3.5 (f) to APC 3.5 (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5061-5311
Cables
BNC,122cm(48in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 10503A SMA,61cm(24in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...8120-1578
Adjustment Procedures 2-37