Procedure 7. A6Al High Voltage Assembly
16.Disconnect ribbon cable A6AlWl from A6J5. See Figure
17.Remove the two screws (1) securing two
18.Remove the two left
19.Lift up the Al7 CRT driver assembly and disconnect A6AlW2 from A17J6. Do not remove any other cables from the Al 7 assembly.
20.Disconnect the tie wraps from the A6Al assembly cables and remove the A6Al High Voltage assembly from the analyzer.
SK136
Figure
Replacement
1.Secure the A6Al High Voltage assembly to the A6 Power Supply using three panhead screws and washers. Connect ribbon cable A6AlWl to A6J5.
2.Snap
3.Place the black grommet protecting the
4.Ensure that all cables are safely routed and will not be damaged when securing the A6 cover.
5.Secure the power supply cover shield to the power supply using three flathead screws (1). See Figure
6.Connect A6AlW2 to A17J6.
7.Place the Al7 CRT Driver assembly into the
8.Secure the Al7 assembly with the two flathead screws removed in “Removal” step 17. See Figure
9.Connect the two mounting posts to the
Assembly Replacement