Table 1-4. Recommended Test Equipment

 

Instrument

 

Critical Specifications

Recommended

Use

 

 

for Equipment Substitution

 

Model

 

 

Synthesized

Frequency Range: 10 MHz to 7 GHz

HP 8340A*

PAT,

 

Sweeper

Frequency Accuracy (CW): 1 x 10eg/day

 

 

M,V

 

(two required)

Leveling Modes: Internal & External

 

 

 

 

 

Modulation Modes: AM & Pulse

 

 

 

 

 

Power Level Range: -80 to +16 dBm

 

 

 

 

Synthesizer/

Frequency Range: 200 Hz to 80 MHz

HP 3335A*

PAT,

 

Level Generator

Frequency Accuracy: 1 x 10e7/month

 

 

M,V

 

 

Flatness: f0.15 dB

 

 

 

 

 

Attenuator Accuracy: <&0.09 dB

 

 

 

 

Synthesized

Frequency Range: 5 MHz to 2.5 GHz

HP 8663A

p,v

 

Signal Generator

Residual SSB Phase Noise at 10 kHz offset

 

 

 

 

 

(320 MHz <fc <640 MHz): <-131 dBc/Hz

 

 

 

 

Pulse/Function

Frequency Range: 10 kHz to 50 MHz

HP 8116A

P

 

Generator

Pulse Width:

200 ns;

 

 

 

 

 

Output Amplitude: 5 V peak-to-peak

 

 

 

 

 

Functions: Pulse & Triangle

 

 

 

 

 

TTL Sync Output

 

 

 

 

AM/FM

Frequency Range: 1 MHz to 200 MHz

HP 8640B

A

 

Signal Generator

Frequency Modulation Mode

 

 

 

 

 

Modulation Oscillator Frequency: 1 kHz

 

 

 

 

 

FM Peak Deviation: 5 kHz

 

 

 

 

Microwave

Frequency Range: 9 MHz to 6.9 GHz

HP 5343A*

P,A,M,V

 

Frequency Counter

Time base Accuracy (Aging): <5 x lo-“/day of

Option 001

 

 

 

External Frequency Reference Input

 

 

 

 

Universal

Modes: TI AFB, Frequency Count

HP

5334A/B

P

 

Counter

Time Interval Measurement Range: 20 ms to 75 s

 

 

 

 

 

Frequency Measurement Range: 17 kHz to 11 MHz

 

 

 

 

 

Frequency Resolution: 1 MHz at 10 MHz

 

 

 

 

 

External Frequency Reference Input

 

 

 

 

 

Time base accuracy (Aging): <3 x 10m7/month

 

 

 

 

 

 

 

 

 

 

Oscilloscope

Bandwidth (3 dB): dc to 100 MHz

HP

1980A/B*

A , T

 

 

Minimum Vertical Deflection Factor: <2 mV/div

 

 

 

 

 

 

 

 

 

Spectrum Analyzer

Frequency Range: 4 kHz to 7 GHz

HP 8566A/B

PAT

 

 

Typical Residual FM: <l Hz Pk-to-Pk in 100 ms

 

 

 

 

 

(Fundamental

Mixing)

 

 

 

 

 

 

 

 

 

 

* Part of Microwave Workstation

P = Performance Tests; A = Adjustments; M = Test & Adjustment Module; T = Troubleshooting; V = Operation Verification

General Information l- 13