14.External Mixer Bias Adjustment
3.On the HP 8560A press [m) EXTERNAL MIXER BIAS, then BIAS OFF.
4.Adjust A15R926 EXT BIAS ZERO for a DVM reading of 0.000 Vdc 512.5 mV.
14.External Mixer Bias Adjustment
3.On the HP 8560A press [m) EXTERNAL MIXER BIAS, then BIAS OFF.
4.Adjust A15R926 EXT BIAS ZERO for a DVM reading of 0.000 Vdc 512.5 mV.