9.Tracking Generator Power Level Adjustments

10.Set the TRK GEN RF POWER to -10dBm. Note the power displayed on the measuring receiver.

Power at -10 dBm Setting

 

dBm

11.Proceed with steps 12 through 14 only if the power level noted in the previous step was outside the range of -10 dBm +/-.23 dB.

12.With the TRK GEN RF POWER set to -10-10 dBm +/-.l dB reading on the measuring location.

dBm, adjust AlOR -10 dB ADJ for a receiver. Refer to Figure 2-12 for adjustment

13.Set the TRK GEN RF POWER to 0 dBm. Adjust AlOR 0 dB ADJ for a 0 dBm

f0.2 dB reading on the measuring receiver. Refer to Figure 2-12 for adjustment location.

14.Repeat steps 12 and 13 until the output power level is within the tolerances indicated at both the -10 dBm and 0 dBm TRK GEN RF POWER settings. Adjust -10 dB ADJ only with the TRK GEN RF POWER set to -10 dBm and adjust 0 dB ADJ only with the TRK GEN RF POWER set to 0 dBm.

2-44 Adjustment Procedures