2. Set the HP 8560A to the following settings:
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OHz C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .389.5MHz CF STEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5 MHz
3.Use the step keys to tune the CENTER FREQ to the values listed in Table
4.As the sampling oscillator frequency is increased, the DVM reading should also increase. The voltage remains at approximately +5.5 Vdc. If the tune voltage is correct, but the ADC measures the voltage and determines it to be out of specification, troubleshoot the A3 assembly’s ADC MUX.
Table
HP 8560A | sampling |
Zenter Frequency (MHz) | Oscillator’s Frequency (MHz) |
|
|
389.5 | 280.0 |
427.0 | 282.5 |
449.5 | 284.0 |
464.5 | 285.0 |
479.5 | 286.0 |
502.0 | 287.5 |
509.5 | 288.0 |
539.5 | 290.0 |
569.5 | 292.0 |
577.0 | 292.5 |
599.5 | 294.0 |
614.5 | 295.0 |
629.5 | 296.0 |
652.0 | 297.5 |
659.5 | 298.0 |
|
|
5.Disconnect W37 from A14J301.
6.Connect a test cable from W37 to the input of another spectrum analyzer. Tune the other spectrum analyzer to the following settings:
Center Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . | 10 MHz |
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. | 2MHz |
7.The amplitude of the 10 MHz reference signal should measure
8.Reconnect W37 from A14J301.
9.Connect the CAL OUTPUT to INPUT 50R.
Synthesizer Section