4. IF Amplitude Adjustments
Procedure
1.Set the HP 8560A m switch to off. Remove the analyzer cover and place the analyzer in the service position as illustrated in Figure
2.Disconnect W29, violet coax cable, from A5J3. Connect the test cable between A5J3 and the 50 R output of the HP 3335A. Set the HP 8560A (LINE) switch on.
3.Set the HP 8560A controls as follows:
C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..10.7MHz SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200kHz REF LVL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.On the HP 8560A, press INIKR) LCAL) and IF ADJ OFF.
5.Set the HP 3335A controls as follows:
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.7MHz AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.Note the marker value. Ideally it should read
7.If the marker reads below
8.If the marker reads above
Note | If A4R826 | has inadequate range, refer to “Inadequate CAL OSC AMPTD |
| Range” in | Chapter 9. |
|
|
|
9.Press AD3 CURR IF STATE . After allowing the analyzer time to complete the adjustments, the displayed amplitude and marker reading should change.
10.Repeat steps 7 and 8 until the marker reads
11.Disconnect the test cable from A5J3 and reconnect W29 to A5J3.