Procedure 5. A2, A3, A4, and A5 Assemblies

SK129

Figure 3-7. A2, A3, A4, and A5 Assembly Removal

Replacement

1.Place the analyzer top-side-up on the work bench.

2.Attach the assembly being installed to the two chassis hinges with two panhead screws.

3.Leave the assembly in the folded-out position and attach ribbon cables Wl and W2.

4.Attach all coaxial cables to the assembly, as illustrated in Figure 3-9.

5.Locate the cable clip on the inside of the right-side frame. Make sure that the coaxial cables are routed properly on the clip as illustrated in Figure 3-9.

6.Lay the A2, A3, A4, and A5 assemblies flat against each other in the folded-out position. Make sure that no cables become pinched between any two assemblies.

3-14 Assembly Replacement