Caution If a dc block is not used, damage to the HP 8566A/B results. The
HP 8566A/B and many other spectrum analyzers have
6.Set the HP 8566A/B to single trigger and press TRACE A
The passband of the filters should be flat within 2 dB over 800 Hz.
7.If continuing to the next procedure, “4.8 kHz and 10.7 MHz IF Filters,” press TRACE A (VIEW) on the HP 8566A/B.
8.Reconnect W27 (coax 3) to A4J3.
4.8 kHz and 10.7 MHz IF Filters
1.Connect a 10.7 MHz signal from a signal generator to the input of the A5 IF assembly and decrease the signal generator level to
2.Fine tune the frequency of the signal generator to center the signal on the screen. Set the signal generator to sweep one 2 kHz span about this center frequency. Set the spectrum analyzer to single sweep.
3.Set the HP 8566A/B to 4.8 kHz center frequency and 2 kHz span.
4.Connect the VIDEO OUTPUT ( rear panel) of the spectrum analyzer through a 20 dB attenuator and dc block to the input of the HP 8566A/B. Set the sweep time of the HP 8566A/B to 10 seconds.
Caution Damage to the HP 8566A/B results if a dc block is not used. The HP 8566A/B and many other’spectrum analyzers have
5.Set the HP 8566A/B to single trigger and press TRACE B
6.To view the bandwidth of the 10.7 MHz IF filters alone, proceed as follows:
a.If the peak of the active trace (trace B) is not at the same amplitude as the stored trace (trace A), adjust the HP 8566A/B reference level as necessary and repeat step 5.
b.Press TRACE B m on the HP 8566.
C. On the HP 8566, press
d. Press (A=), DISPLAY LINE [a’). Adjust the display line such that it is at the peak of trace B (the dimmer trace). Press
e. On the HP 8566A/B, press [FREQUENCY SPAN), 0 IHz),
(SINGLE).