Removal
Warning The A6 Power Supply and A6Al High Voltage assemblies contain lethal voltages with lethal currents in all areas. Use extreme care when servicing these assemblies. Always disconnect the power cord from the instrument before beginning this replacement procedure. Failure to follow this precaution can represent a shock hazard which may result in personal injury.
1.Disconnect the power cord from the analyzer.
2.Remove the analyzer’s cover assembly as described in Procedure 1.
3.Fold out the A2, A3, A4, and A5 assemblies as described in Procedure 5, “A2, A3, A4, and A5 Assemblies.”
4.Place the analyzer
Warning The voltage potential at A6AlW3 is +9 kV. Disconnect at the CRT with caution! Failure to properly discharge A6AlW3 may result in severe electrical shock to personnel and damage to the instrument.
5.Connect the analyzer’s
6.Connect a
7.Connect the clip lead of the probe (ground) to the chassis of the spectrum analyzer.
8.Slip the tip of the
9.Keep the
10.Disconnect the
11.Using a small screwdriver with the shank in contact with the CRT shield assembly, slip the tip of the screwdriver under the A6AlW3
12.Pry out the black grommet protecting
13.Carefully unsnap the A6AlW3
14.Remove the three screws securing the power supply shield to the power supply and remove the shield.
15.Remove the three screws and washers securing the A6Al High Voltage assembly to the A6 Power Supply assembly.