RE F/ I”.; dBmi ATTEN 20 dB
5dB,’
S A M P L E
1 I I I I
C E N T E R 1 0 . 7 0 0 0 0 0 | MHz |
| S P A N 0 H z |
R E S B W 3 0 0 | kHz |
| |
V B W 3 0 0 kHz | S W P 5 0 0 set |
S K 1 8 1
Figure
1 MHz Resolution Bandwidth Problems
Check the crystal shorting switches as follows:
1. On the spectrum analyzer, press [PRESET) and set the controls as follows:
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . l M H z 500kHz
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..300MH z
2.On the spectrum analyzer, connect the 300 MHz CAL OUTPUT to the INPUT 50ohm.
3.If the trace flatness is not within 2.5 dB, a failure probably exists.
4.A trace similar to Figure
5.Set the spectrum analyzer (SPAN) to 3 MHz. A trace that slopes across the screen (see Figure