Table
i | - |
|
|
|
|
|
| ||
| Detector Mode | U108 pin 1 | U108 pin 15 | U108 pin 16 |
|
|
|
|
|
| SAMPLE | H | L | H |
| POS PEAK | H | L | L |
| NEG PEAK | L | H | H |
|
|
|
|
|
5. Set the HP 8560A to the following settings:
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..lMHz SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50ms
6.Check for the presence of the SCAN RAMP signal by connecting an oscilloscope probe to A3J2 pin 45 ( component side of A3J2). Connect the
7.A 0 to 10 V ramp should be present in both LINE and FREE RUN trigger modes. If the waveform is present only in LINE trigger, ADC control signal
8.If the scan ramp is present, but is not being switched to the output of U108, replace
U108. If the scan ramp is absent in either mode, do the following:
a.Connect the oscilloscope probe to A3J400 pin 15 (HSCAN).
b.A TTL signal (high during 50 ms sweep time and low during retrace) should be present, indicating A3 is working properly. Refer to the Synthesizer Section troubleshooting procedure. A faulty TTL signal indicates a bad A3 Interface assembly.
9.Set the HP 8560A to the following settings:
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100ms
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..lOOMHz
IF Adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ON
10.Check for the presence of the CAL OSC TUNE signal by monitoring A3J401 pin 25 with an oscilloscope. If ERR 499 CAL UNLK is displayed and a signal greater than 10 V
11.If a constant dc voltage is present during the sweep and all of the retrace period, refer to the IF Section troubleshooting procedure.