Procedure 14. W3 Line Switch Cable

21.From the top side of the analyzer, use contact removal tool, HP part number 8710-1791, to remove the four wires from the W3 connector. See Figure 3-26. With wire cutters, clip the tie wrap holding the cable to the contact housing.

22.Completely remove the cable from the instrument.

23.Remove the Al Front-Frame assembly and Al8 CRT assembly as described in Procedure 2, “Al Front Frame/Al8 CRT Removal,” steps 16 through 29.

24.Remove the left side frame from the analyzer using the hardware listed below. (The side frame will still be attached by the speaker wires. Do not let it hang freely.)

Screw

(1)SCREW-MACH M4 X 0.7 8 mm-LG FLAT HD

(2)SCREW-MACH M3 X 0.5 35 mm-LG FLAT HD (3)SCREW-MACHM3X0.56mm-LGFLATHD

25.Remove the line-switch cable assembly.

Quantity

. . . . . . . . . . . . . . . . . . . . . . .3

. . . . . . . . . . . . . . . . . . . . . .2

. . . . . . . . . . . . . . . . . . . . . . . 6

D R E S S C A B L E B E T W E E N

S T A N D O F F A N D S I D E F R A M E

Figure 3-25. Line Switch Mounting Screw and Cable Dress

3-46 Assembly Replacement