Procedure 2. Al Front Frame/Al8 CRT

Warning The voltage potential at A6AlW3 is +9 kV. Failure to discharge A6AlW3 correctly may result in severe electrical shock to personnel and damage to the instrument.

A6AlW3

HIGH-VOLTAGE PROBE

G R O MME T\ A6A-1 W3

SK123

Figure 3-2. Discharging the Post-Accelerator Cable