6. Sampling Oscillator Adjustment
6. Sampling Oscillator AdjustmentAssembly Adjusted
Al5 RF Assembly
Related Performance Test
There is no related performance test for this adjustment procedure.
Description
The phase detector bias is adjusted for 1.8 Vdc. The sampling oscillator tank circuit is adjusted for a tuning voltage of 0.9 Vdc when the oscillator is set to 288 MHz. The voltage monitored is actually the tuning voltage divided by 4.05. A
| SPECTRUM ANALYZER | |
D I G I T A L V O L T M E T E R | w PC | |
BOARD | ||
| ||
| P R O P |
S K 1 7
Figure
Equipment
An asterisk (*) indicates equipment used only for the
Digital Voltmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3456A
DVM Test Leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 34118A
Frequency Counter* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 5343A
Active Probe* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 1121A
Power Supply* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 6114A 1 MR Resistor* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adapters
Type BNC (f) to BNC (f)’ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cables
BNC,122cm(48in)* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 10503A
Adjustment Procedures