15. External Mixer Amplitude Adjustment
Adapters
Type N (f) to SMA
Cables
BNC,122cm (48in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 10503A SMA,61cm
Procedure
1.Set the HP 8560A (LINE] switch off and disconnect the power cord. Remove the analyzer cover and reconnect the power cord.
2.Set up the equipment as illustrated in Figure
3.Move the WR PROT/WR ENA jumper on the A2 Controller assembly to the WR ENA position. Set the HP 8560A luNE) switch on.
4.On the HP 8560A, press (AUX) EXTERNAL MIXER AMPTD CORRECT , then
CNV LOSS VS FREQ .
5. On the HP 8560A, press LCAL), MORE 1 OF 2, SERVICE CAL DATA, 3RD IF AMP, and
CAL 3RD AMP GAIN. Wait until the message
Note | Perform the following steps only if the HP 8560A does not have Option 002. |
6.Press @ or @ to display the conversion loss value for each frequency listed in Table
7.If all conversion loss values equal 30 dB, skip to step 8, otherwise continue to step a.
a.Refer to Table
b.Use the analyzer
c.Repeat steps a and b for all frequencies having a conversion loss value other than 30 dB.
8.Press [INSTR PRESET) on the HP 8340A/B and set the controls as follows:
CW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 310.7 MHz POWERLEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .