Table
| Instrument | Critical Specifications | Recommended | Use | |||||
|
|
|
| for Equipment Substitution | Model |
|
| ||
| 20 dB Fixed | Frequency Range: dc to 6.5 GHz | HP 8491B | p,v | |||||
| Attenuator | Attenuation Accuracy: <fl dB | Option 020 |
|
| ||||
|
|
|
| Maximum SWR: 1.2 (dc to 6.5 GHz) |
|
|
| ||
|
|
|
|
|
| ||||
| 10 dB Fixed | Frequency Range: dc to 6.5 GHz | HP 8491B | p,v | |||||
| Attenuator | Attenuation Accuracy: <f0.6 dB | Option 010 |
|
| ||||
|
|
|
| Maximum SWR: 1.2 (dc to 6.5 GHz) |
|
|
| ||
| Signature | Multimeter | Clock Frequency >lO MHz | HP 5005A/B | T | ||||
| Reference | Attenuator | Supplied with | HP | 8484A | HP 11708A | P,A | ||
| Termination | Frequency Range: dc to 22 GHz | HP 909D | P,M,V | |||||
|
|
|
| Impedance: 5OQ |
|
|
|
| |
|
|
|
| Maximum SWR: <1.22 |
|
|
| ||
|
|
|
| Connector: APC 3.5 |
|
|
| ||
| Filter | Cutoff Frequency: | 50 MHz | P,M,V | |||||
|
|
|
| Rejection at 80 MHz: >50 dB |
|
|
| ||
| 12 MHz | Cutoff Frequency: 12 MHz | P | ||||||
|
|
|
|
|
|
| |||
|
|
|
| Rejection at 18 MHz: >45 dB |
|
|
| ||
|
|
|
|
|
|
| |||
| Directional |
| Frequency Range: 5 MHz to 50 MHz | HP 8721A | P | ||||
| Bridge |
| Coupling: 6.0 | dB | (nominal) |
|
|
| |
|
|
|
| Directivity: 30 | dB | minimum |
|
|
|
|
|
|
| VSWR: <1.45 |
|
|
|
|
|
| Power Splitter | Frequency Range: 1 kHz to 22 GHz | HP 11667B | P,A,M,V | |||||
|
|
|
| Insertion Loss: 6 dB (nominal) |
|
|
| ||
|
|
|
| Output Tracking: <0.25 dB |
|
|
| ||
|
|
|
| Equivalent Output | SWR: <1.22 |
|
|
| |
| Product | Support Kit | No Substitute |
|
| A,T | |||
| Adapter |
|
| Type N (f) to BNC (m) | A | ||||
|
|
|
|
|
|
|
| ||
| Adapter |
|
| Type N (m) to BNC (f) | p,v | ||||
| (three | required) |
|
|
|
|
|
| |
| Adapter |
|
| Type N (f) to BNC (f) | 1250- 1474 | p,v |
| ||
| Adapter |
|
| Type N (f) to APC 3.5 (m) | A |
| |||
| Adapter |
|
| Type N (m) to APC 3.5 (m) | P,A,M,V | ||||
| (two | required) |
|
|
|
|
|
| |
|
|
|
|
|
|
|
| ||
| Adapter |
|
| Type N (m) to APC 3.5 (f) | P,A,V |
|
P = Performance Tests; A = Adjustments; M = Test & Adjustment Module; T = Troubleshooting; V = Operation Verification
General Information