The EEROM on A2 is used to store data for frequency response correction, elapsed time, focus, and intensity levels. Error codes from 700 to 703 indicate that some part of the data in EEROM is invalid. An EEROM error could result from either a defective EEROM or an improper sequence of storing data in EEROM. Check the EEROM with the following steps:
1.Place the WR PROT/WR ENA jumper on the A2 Controller assembly in the WR ENA position.
2.On the spectrum analyzer, press ICAL) MORE 1 OF 2, SERVICE CAL DATA, FLATNESS ,and FLATNESS DATA. Enter a value of 130. Press PREV MENU, STORE DATA , YES, and [m).
3.Press INTEN, enter an intensity value of 90, and press
STORE INTEN.
4.Press MORE 1 OF 2 FOCUS, enter a focus value of 128, and press
STORE FOCUS. Turn the (LINE) switch off, then on, cycling the power.
5.If errors are still present, the EEROM A2U501 is defective. Refer to the EEROM replacement procedure in Chapter 3.
700 EEROM
701 AMPL CAL
702 ELAP TIM
703 AMPL CAL
704 PRESELCT
Program ROM
| The instrument |
| programmed ROM (A2 Controller assembly). If an invalid |
| is found for a particular ROM, an error code is generated. If a |
| defective programmed ROM is found, replace it with another ROM |
| with the same HP part number. Refer to Chapter 4, “Replaceable |
| Parts.” |
| Although some of these errors might result in a blanked display, it is |
| possible to read these errors over |
| a Functional Section” in this chapter. |
705 ROM U306 | |
706 ROM U307 | |
707 ROM U308 |