All YTO

Removal

1.If the analyzer is an Option 002, remove the Tracking Generator before proceeding.

2.Place the analyzer top-side-down on the work bench.

3.Use a 5/16 inch wrench to remove W56/FL2/W57 (as a unit).

4.Disconnect W38 at the All Assembly.

5.Remove the All mounting screws (1) shown in Figure 3-18.

6.Disconnect WlO from All.

SJ19

Figure 3-18. A11 Mounting Screws

Replacement

1.Reconnect WlO to All.

2.Place the All Assembly in the analyzer.

3.Secure the All Assembly to the right-side frame using the four screws (1) removed in “Removal” step 5.

4.Connect W38 to All.

5.Install W56/FL2/W57. Ensure that all of the connections are tight.

6.If the analyzer is an Option 002, install the Tracking Generator.

3-32 Assembly Replacement