Removal
Warning Due to possible contact with high voltages, disconnect the analyzer’s
1.Remove the analyzer’s cover assembly as described in Procedure 1, “Analyzer Cover.”
2.Fold out the A2, A3, A4, and A5 assemblies as described in Procedure 5, “A2, A3, A4, and A5 Assemblies Removal,” steps 3 through 5.
3.Disconnect AlAlWl from A3J602.
4.Place the analyzer
Warning The voltage potential at A6AlW3 is +9 kV. Disconnect at the CRT with caution! Failure to properly discharge A6AlW3 may result in severe electrical shoclc to personnel and damage to the instrument.
5.Connect the analyzer’s
6.Connect a
7.Connect the clip lead of the probe (ground) to the chassis of the spectrum analyzer.
8.Slip the tip of the
9.Keep the
10.Disconnect the
11.Using a small screwdriver with the shank in contact with the CRT shield assembly, slip the tip of the screwdriver under the A6AlW3
12.Remove the three screws securing the power supply shield to the power supply, and remove the shield.
13.Pull the cable tie (l), Figure
14.Disconnect W3 from A6J2.
15.Pull W3 up from between the power supply and the CRT assembly.
16.Undress W3.
17.Place the analyzer on its right side frame.