8.If the reset pulses are absent, troubleshoot the Peak Detector Reset circuitry.
9.If the reset pulses are present, check the gate of Q207. The pulses should be
10.The peak detector can be made into a unity gain amplifier by shorting the cathode of CR203 to the anode of CR204. If the peak detector functions normally as a unity gain amplifier, suspect Q208 or CR203 or CR204.
Peak Detector Reset
See function block R of A3 Interface Assembly Schematic Diagram (sheet 4 of 6).
1. Press (PRESET] on the HP 8560A and set the controls as follows:
C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...300MHz
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Hz
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5s
DETECTOR MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . POS PEAK
2.Check that HHOLD (A3U526 pin 11) has 18 ps wide pulses every 128 pus.
3.Check that HODD (U408 pin 5) is a square wave with a period of 16.7 ms (2 x sweep time/600).
4.Check
5.Check LNEG_RST (A3U422 pin 12) for 200 ns
6.Set the detector mode to NORMAL and check that LNEG_RST (A3U422 pin 12) has two pulses spaced 40 pus apart and then a single pulse approximately 88 ps from the second pulse.
7.Check HMUXSELO (A3U408 pin 3) and HMUXSELl (A3U408 pin 9) according to Table
Table
Detector Mode | ||
|
|
|
NORMAL | 40 ps pulse every 128 ps | 40 ps pulse every 128 ps |
SAMPLE | H | H |
POS PEAK | H | L |
NEG PEAK | L | H |
|
|
|
Rosenfell Detector
See function block S of A3 Interface Assembly Schematic Diagram (sheet 4 of 6).
If both
HROSENFELL will also be set high. This indicates that the video signal probably consists of noise, since it rose and fell during the same period. The HROSENFELL signal is valid only when the NORMAL ( rosenfell) detector mode is selected.