8.If the reset pulses are absent, troubleshoot the Peak Detector Reset circuitry.

9.If the reset pulses are present, check the gate of Q207. The pulses should be positive-going from -12.7 V to -1.35 V.

10.The peak detector can be made into a unity gain amplifier by shorting the cathode of CR203 to the anode of CR204. If the peak detector functions normally as a unity gain amplifier, suspect Q208 or CR203 or CR204.

Peak Detector Reset

See function block R of A3 Interface Assembly Schematic Diagram (sheet 4 of 6).

1. Press (PRESET] on the HP 8560A and set the controls as follows:

C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...300MHz

SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Hz

SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5s

DETECTOR MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . POS PEAK

2.Check that HHOLD (A3U526 pin 11) has 18 ps wide pulses every 128 pus.

3.Check that HODD (U408 pin 5) is a square wave with a period of 16.7 ms (2 x sweep time/600).

4.Check LPOS-RST (U422 pin 4) for 200 ns low-going pulses every 128 pus.

5.Check LNEG_RST (A3U422 pin 12) for 200 ns low-going pulses every 128 ps.

6.Set the detector mode to NORMAL and check that LNEG_RST (A3U422 pin 12) has two pulses spaced 40 pus apart and then a single pulse approximately 88 ps from the second pulse.

7.Check HMUXSELO (A3U408 pin 3) and HMUXSELl (A3U408 pin 9) according to Table 7-8.

Table 7-8. HMUX-SELO/l Versus Detector Mode

Detector Mode

HMUX-SELO

HMUX-SELl

 

 

 

NORMAL

40 ps pulse every 128 ps

40 ps pulse every 128 ps

SAMPLE

H

H

POS PEAK

H

L

NEG PEAK

L

H

 

 

 

Rosenfell Detector

See function block S of A3 Interface Assembly Schematic Diagram (sheet 4 of 6).

If both HPOS-HLDNG and HNEG-HLDNG are high during the same bucket,

HROSENFELL will also be set high. This indicates that the video signal probably consists of noise, since it rose and fell during the same period. The HROSENFELL signal is valid only when the NORMAL ( rosenfell) detector mode is selected.

7-18 ADC/lnterface Section