564 LOG AMPL Unable to adjust amplitude of log scale. No compression in CalLogPower .

565 LOG AMPL Unable to adjust amplitude of log scale. Gain compression error in CalLogPower.

566 LOG AMPL Unable to adjust amplitude of log scale. Unable to set LOG CAL LVL in CalLogPower

567 LOG AMPL Unable to adjust amplitude of log scale. No TOS LX in CalLogExpand. Possible problem in Log offset/Log Expand stage.

568 LOG AMPL Unable to adjust amplitude of log scale. No LVL LX in CalLog Power. Possible problem in Log offset/Log Expand stage.

569 LOG AMPL Unable to adjust amplitude of log scale. Low MDACX value in CalLogExpand. Possible problem in Log offset/Log Expand stage.

570 LOG AMPL Unable to adjust amplitude of log scale. High MDACX value in CalLogExpand. Possible problem in Log offset/Log Expand stage.

571 AMPL Unable to adjust step gain amplifiers. No TOS-D in CalDetectors.

572 AMPL IMUnable to adjust amplitude of 1 MHz Resolution Bandwidth. No TOS-W in CalWidebandLog.

573 LOG AMPL Unable to adjust amplitude of log scale. Video offset error +O. Check video offset circuitry.

574 LOG AMPL

575 LOG AMPL

Unable to adjust amplitude of log scale. Video offset error t.1. Check video offset circuitry.

Unable to adjust amplitude of log scale. Video offset error +2. Check video offset circuitry.

576

LOG

AMPL

Unable to adjust amplitude of log scale. Video offset error +3. Check

 

 

 

video offset circuitry.

577

LOG

AMPL

Unable to adjust amplitude of log scale. Video offset error +4. Check

 

 

 

video offset circuitry.

578

LOG

AMPL

Lim Calibration error from DC Logger Cal

579

LOG

AMPL

Attenuator CAL level error from DC Logger Cal.

580

LOG

AMPL

FID CAL level error from DC Logger Cal

581

AMPL

 

Unable to adjust 100 kHz and 510 kHz Resolution Bandwidths.

 

 

 

ADC/CALOSC handshake CAL in Sweep Xtal problem. Refer to

 

 

 

Error 582.

General Troubleshooting 6-29