Procedure 1. Analyzer Cover

Procedure 1. Analyzer Cover

Removal/Replacement

1.Disconnect the line-power cord, and place the analyzer on its front panel.

2.Loosen (but do not remove) the four rear-bumper screws, using a 4 mm hex wrench. Pull the cover assembly off towards the rear of the instrument.

Caution When replacing the analyzer’s cover, use caution to avoid damaging any cables.

3.When installing the cover assembly, be sure to locate the cover’s air vent holes on the bottom side of the analyzer. Attach with the four screws loosened in step 2, and tighten the four screws gradually to ensure that the cover is seated in the front-frame gasket groove.

4.Torque each screw to 40 to 50 inch-pounds to ensure proper EM1 gasket compression.

SK122

Figure 3-1. Hinged Assemblies

Assembly Replacement 3-3