582 AMPL Unable to adjust 100 kHz and 210 kHz Resolution Bandwidths. Bad CALOSC Calibration in Sweep Rate. Test the 100 kHz Resolution Bandwidth filter 3 dB bandwidth as follows:
1.Connect the CAL OUTPUT signal (A4J8) to the INPUT 50R.
2.Press (PRESET) and set the controls as follows:
CENTER FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . .300 MHz
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500kHz
Resolution Bandwidth . . . . . . . . . . . . . . . . . . . . . . 100 kHz
LOG dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 dB
REF LEVEL . . . . adjust to place signal peak at top of the screen
3.Press (PEAK SEARCH) and MARKER DELTA and turn the knob clockwise to position the marker until the delta MKR reads
+/
4.Press MARKER DELTA and move the marker to the other side of the peak until the delta MKR reads 0 dB
5.If the delta MKR frequency is between 90 kHz and 110 kHz, the 100 kHz Resolution Bandwidth is working properly. If the frequency is outside these limits, read the following information on the A4 Cal Oscillator sweep generator.
If the 100 kHz Resolution Bandwidth works properly, the A4 Cal Oscillator sweep generator is failing to sweep its oscillator frequency at the correct rate. The error is detected in sweeping on the skirts of the 100 kHz Resolution Bandwidth.
A properly operating sweep generator generates a series of
1.Remove the shields.
2.Connect an oscilloscope probe to A4U804C pin 8.
3.On the spectrum analyzer, press (CAL) and FULL IF ADJ .
4.Approximately 8 seconds after starting the FULL IF ADJ , check for
583 RBW 30K Unable to adjust 30 kHz Resolution Bandwidth. SYSBW and LCBW disagreement in LCCAL.
584 RBW IOOK Unable to adjust 100 kHz Resolution Bandwidth. SYSBW and LCBW disagreement in LCCAL.