5. DC Log Amplifier Adjustments
Procedure
1.Set the HP 8560A m switch to off. Remove the analyzer cover and place the analyzer in the service position as illustrated in Figure
2.Connect the HP 3335 50 0 output to the HP 8560A 50 0 input. Set the HP 8560A (LINE) switch on.
3.Set the HP 8560A controls as follows
C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...15 MHz SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..O REF LVL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. Set up an HP 3335A as follows:
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...15 MHz AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.Press ICAL), ADJ CURR IF STATE , wait for the analyzer to complete adjustments then press m.
6.Adjust A4R445 for maximum
A4 Linear Fidelity Adjustment
This adjustment consists of adjusting A4R544 until the delta marker reads
Procedure
1.Set the HP 8560A (LINE) switch to off. Remove the analyzer cover and place the analyzer in the service position as illustrated in Figure
2.Connect the HP 3335 50 R output to the HP 8560A 50 R input. Set the HP 8560A m switch on.
3.Press(PRESET)(AMPLITUDE), LINEAR, MORE 1 of 3, AMPD UNITS,dBm,(CAL),
IF ADJ OFF.
4. Set the HP 8560A controls as follows:
C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...15 MHz
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...5 MHz
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300kHz
REF LVL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5. Set up an HP 3335A as follows:
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...15 MHz AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6. Press [PEAK SEARCH), MARKER DELTA .