17. Signal ID Oscillator Adjustment

9.While observing the HP 8566A/B display, adjust A15C629 SIG ID for the highest obtainable frequency with less than 3 dB decrease in amplitude from maximum. Read this frequency from the frequency counter and record as Fs dB HIGH.

F3 dB HIGH =

 

Hz

 

10.Observe the HP 8566A/B display as you adjust A15C629 SIG ID for the lowest obtainable frequency with less than 3 dB decrease in amplitude from maximum. Record the frequency counter reading as Fs dB LOW.

F3 dB L O W =

 

MHz

 

11.Calculate the difference between Fs dB HIGH and Fs dB LOW, then divide results by four. Enter the result as FOFFSET.

FOFFSET =

kHz

12. Add FOFFSET to Fs dB LOW recorded in step 10 and record the result as FSIGID.

F SIGID =

 

MHz

 

13.Adjust A15C629 for a frequency counter reading equaling FSIGID. The final adjusted frequency must equal 10.7 MHz f50 kHz.

Adjustment Procedures 2-65