TROUBLESHOOTING AND REPAIR

F-41

AUXILIARY TRANSFORMER NO. 1 TEST (continued)

 

TABLE F.3 — SECONDARY VOLTAGES.

 

 

 

 

 

NORMAL EXPECTED

 

LEAD IDENTIFICATION

VOLTAGE

 

 

 

 

X1 to X2

52 VAC

 

X3 to X5

115 VAC

 

X3 to X4

24 VAC

 

 

 

 

Return to Section TOC

Return to Section TOC

Return to Section TOC

Return to Master TOC

Return to Master TOC

Return to Master TOC

7.Carefully apply the correct input voltage to the Power Wave 455 and check for the correct secondary voltages per Table F.3.

NOTE: The secondary voltages will vary if the input line voltage varies.

8.If the correct secondary voltages are present, the T1 auxiliary transformer is functioning properly. If any of the sec- ondary voltages are missing or low, check to make certain the primary is configured correctly for the input volt- age applied. See wiring diagram.

WARNING

High voltage is present at primary of Auxiliary Transformer.

9.If the correct input voltage is applied to the primary, and the secondary volt- age(s) are not correct, the T1 trans- former may be faulty.

10.Install the case sides and top using the 3/8 in. nutdriver.