Chapter 2 103
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
Alpha.
The filter is unavailable when the measurement method is set to RBW.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: Off
History: Added with PSA firmware revision A.03.00.
Added with ESA firmware revision A.08.00.
Remote Command:
[:SENSe]:ACPower:FILTer[:RRC][:STATe] OFF|ON|0|1
[:SENSe]:ACPower:FILTer[:RRC][:STATe]?
Example: ACP:FILT 1
ACP:FILT?
2.3.12 Filter Alpha
Press Filter Alpha to input the alpha value for the RRC Filter. This parameter is available
when 3GPP W-CDMA, NADC, or TETRA has been selected as the Radio Std. from the Mode
Setup menu.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 0.22 when W-CDMA is selected, 0.35 when NADC or TETRA is selected,
otherwise Off.
Knob Increment: 0.01
Step Key
Increment: 0.1
Range: 0.01 to 1.0
History: Added with PSA firmware revision A.03.00.
Added with ESA firmware revision A.09.00.
Remote Command:
[:SENSe]:ACPower:FILTer[:RRC]:ALPHA <number>
[:SENSe]:ACPower:FILTer[:RRC]:ALPHA?
Example: ACP:FILT:ALPH 0.22
ACP:FILT:ALPH?