170 Chapter 2
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
referenced to the last measured value, or you may specify the reference for the
multi-carrier power measurement. Relative values are displayed, referenced to the “Power
Reference” value.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: Off/Measured power in the reference carrier.
Ter mina tors : dB m
Default Terminator: dBm
Range: –200 dBm to 200 dBm
History: Added with PSA firmware revision A.02.00.
Added with ESA firmware revision A.09.00.
Remote Command:
[:SENSe]:MCPower:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:MCPower:CARRier:AUTO[:STATe]?
[:SENSe]:MCPower:CARRier[:POWer]<dBm>
[:SENSe]:MCPower:CARRier[:POWer]?
Example: MCP:CARR:AUTO 0
MCP:CARR:AUTO?
MCP:CARR -100
MCP:CARR?
2.16.9 Limit Test
Pressing Limit Test turns the testing of the limit line on or off. When Limit Test is set to On,
each offset is compared to its upper and lower offset limit. In those cases where the power
exceeds the limit, a red “F” is placed next to the dBc result to indicate a failure. If there are
any failures, the pass/fail indicator area show “FAIL” in red; if there are none, it shows
PASS” in green. Any offsets that are in the off state (see “Offset Freq” on page 165) are not
measured and their results will not be displayed on screen.
Key Path: Meas Setup, More
State Saved: Saved in instrument state.
Factory Preset: Off
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.08.00