Chapter 2 113
One-Button Measurement Functions
Meas Setup (Burst Power)
One-Button Measurement Functions
width. In either case, the burst width is centered on the observed burst in each
measurement.
NOTE This key will be grayed out if Meas Method is set to Above Threshold Lvl.
Key Path: Meas Setup
Factory Preset: 542.77 µs
Terminators: %, s, ms, µs, ns (% from front panel only)
Default Terminator: seconds (s)
Knob Increment: 0.1 µs
Step Key
Increment: 10 µs
Range: 0.1 µs to 2 ks
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00
Remote Command:
[:SENSe]:BPOWer:BURSt:WIDTh <time>
[:SENSe]:BPOWer:BURSt:WIDTh?
[:SENSe]:BPOWer:BURSt:AUTO OFF|ON|0|1
[:SENSe]:BPOWer:BURSt:AUTO?
Remote Command Notes: Burst width cannot be set remotely as a percent of the current
burst width. That functionality is available from the front panel, but the
command must be sent with a time value.
Example: SENS:BPOW:BURST:AUTO ON
2.5.7 Optimize Ref Level
Sets the input attenuator and reference level to optimize the robustness of the
measurement, which is its freedom from errors due to input compression and log amp
range limitations.
NOTE There will always be a minimum of 2 dB for PSA or 5 dB for ESA of
attenuation set to protect the analyzer input.
Key Path: Meas Setup