104 Chapter 2
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
2.3.13 Noise Correction
Pressing Noise Correction turns noise correction on or off. When you set Noise Corr to On, a
calibration of the noise floor is performed and used to correct for analyzer noise floor
contribution to measurement levels, increasing dynamic range.
Noise Correction is unavailable if signal tracking is on, or when the radio standard is set to
NADC or PDC when the measurement method is set to RBW.
Key Path: Meas Setup,
State Saved: Saved in instrument state.
Factory Preset: Off
History: Added with PSA firmware revision A.02.00.
Added with ESA firmware revision A.07.01.
Remote Command:
[:SENSe]:ACPower:CORRection:NOISe[:AUTO] OFF|ON|0|1
[:SENSe]:ACPower:CORRection:NOISe[:AUTO]?
Remote Command Notes: The noise correction feature is not available when the radio
standard is set to NADC or PDC
Example: ACP:CORR:NOIS 1
ACP:CORR:NOIS?