Chapter 2 169
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
attenuation set to reduce the potential errors due to VSWR.
Key Path: Meas Setup
State Saved: Not saved.
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.08.00
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example: POW:RANG:AUTO ONCE
2.16.7 Method
Enables you to set the measurement method to either the integration bandwidth method
(IBW) or the resolution bandwidth method (RBW). The resolution bandwidth method is
most useful for measuring cdmaOne and cdma2000 signals; the integration bandwidth
method is preferred for other signals.
Key Path: Meas Setup
Key Notes: When Method is set to RBW, neither Noise Correction nor RRC Filter are
available.
State Saved: Saved in instrument state.
Factory Preset: IBW
Range: IBW|RBW
History: Added with PSA firmware revision A.02.00.
Added with ESA firmware revision A.09.00.
Remote Command:
[:SENSe]:MCPower:METHod IBW|RBW
[:SENSe]:MCPower:METHod?
Example: MCP:METH RBW
MCP:METH?
2.16.8 Power Ref
Enables you to set the multi-carrier power reference to automatic or manual. When set to
automatic, the carrier power result reflects the measured power value in the selected
reference carrier (Meas Setup, Carrier Setup, Ref Carrier). When set to manual, the result is