200 Chapter 2
One-Button Measurement Functions
Meas Setup (Spectrum Emissions Mask—SEM)
One-Button Measurement Functions
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST
ABSolute|AND|OR|RELative,ABSolute|AND|OR|RELative,ABSolute|AND|OR|RELative
,ABSolute|AND|OR|RELative
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST?
Remote Command Notes: Comma separated list. When offset F is off, this will be 5
values. When offset F is on, this will be 6 values. n = 1 for BTS and n = 2
for MS. Default is BTS.
Example: SEM:OFFS:LIST:TEST ABS,REL,ABS AND REL,ABS OR REL
SEM:OFFS:LIST:TEST?
2.19.4.11.1 Absolute
Set the limit test to show Fail if one of the absolute spectrum emission mask measurement
results is larger than the limit for Abs Start and/or Abs Stop. This is the default selection
for each offset
Key Path: Meas Setup, Offset/Limits, More, Fail Mask
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits” on page 187 for this value.
Remote Command:
Refer to “Fail Mask” on page 199.
2.19.4.11.2 Relative
Sets the limit test to show Fail if one of the relative spectrum emission mask measurement
results is larger than the limit for Rel Start and/or Rel Stop.
Key Path: Meas Setup, Offset/Limits, More, Fail Mask
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits” on page 187 for this value.
Remote Command:
Refer to “Fail Mask” on page 199.
2.19.4.11.3 Abs AND Rel
Set the limit test to show Fail if one of the absolute spectrum emission mask measurement
results is larger than the limit for Abs Start and Abs Stop AND one of the relative spectrum