Chapter 2 127
One-Button Measurement Functions
Marker (Complementary Cumulative Distribution Function—CCDF)
One-Button Measurement Functions
History: Added with ESA firmware revision A.07.00
Remote Command:
See Volume 1, Core Spectrum Analyzer Functions User’s and Programmer’s Reference
guide.
Example: CALC:PST:MARK3:MODE DELT selects marker 3 as a delta marker and
places a reference marker at the 0 dB point on the specified trace (see
“Marker Trace” on page 127). If marker 3 is OFF it places both the active
and the reference markers at the 0 dB point of the specified trace.
2.9.4 Off
Turns off the selected marker. In addition, Off also turns off functions related to the
selected marker.
Key Path: Marker (after MEASURE, Power Stat CCDF is selected.)
History: Added with ESA firmware revision A.07.00
Remote Command:
See Volume 1, Core Spectrum Analyzer Functions User’s and Programmer’s Reference
guide.
Example:
CALC:PST:MARK3:STAT OFF selects marker 3 and sets it to Off.
2.9.5 Marker Trace
Chooses which trace the selected marker will be placed on. You can pick Measured,
Gaussian, or Reference.
Key Path: Marker (after MEASURE, Power Stat CCDF is selected.)
State Saved: The Marker Trace for each marker is saved in instrument state.
Factory Preset: Measured
History: Added with ESA firmware revision A.07.00
Remote Command:
:CALCulate:PSTatistic:MARKer[1]|2|3|4:TRACe MEASured|GAUSsian|REFerence
Puts the marker on the specified trace.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:TRACe?
The query returns the type of the trace on which the marker currently resides.