Chapter 2 215
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
indicate this range is currently inactive.
Key Path: Meas Setup
Saved State: All values for all ranges are saved in instrument state.
Factory Preset: For each range, the following settings are selected: .
Video BW (Auto),
Sweep Time (Auto),
Absolute Limit = 50 dBm,
Peak Threshold = 90 dBm,
Peak Excursion = 6 dB, and
Abs Stop Limit Mode (Auto) is set to 50 dBm
Attenuation (Auto)
Detector 1 = Peak
Detector 2 = Off
Points = (Man) 601
The table below defines the other parameter’s settings for the
measurement. The values for ranges 1 through 5 are based on the
W-CDMA (3GPP) BTS Spurious Emission measurement.
Table 2-6 Range Table Default Settings
Range Start Freq (GHz) Stop Freq (GHz) Res BW (kHz)
1 1.920 1.980 1200
2 1.894 1.920 510