128 Chapter 2
One-Button Measurement Functions
Marker (Complementary Cumulative Distribution Function—CCDF)
One-Button Measurement Functions
Example: CALC:PST:MARK1:TRAC GAUSsian places marker 1 on the Gaussian trace.
2.9.6 Marker All Off
Turns off all markers.
NOTE Selecting any measurement (including Meas Off) under Measure, turns off the
marker table.
Key Path: Marker (after MEASURE, Power Stat CCDF is selected.)
History: Added with ESA firmware revision A.07.00
Remote Command:
:CALCulate:PSTatistic:MARKer[1]|2|3|4:AOFF
Example: CALC:PST:MARK:AOFF turns off all markers.
2.9.7 Marker X Position (Remote Command Only)
Sets the marker X position (horizontal) to a specified point on the X axis in the current
X-axis units (dB). If the value chosen would place the marker off screen, the marker will be
placed at the left or right side of the display, on the specified trace (see “Marker Trace” on
page 127). If the marker is off, this command has no effect.
Key Path: Marker
Key Notes: When a marker is active, you may use the knob, numeric keypad, or ⇑ ⇓
keys to specify the horizontal location of the marker.
Ter mina tors : dB
Step Key
Increment: 1 dB
Range: 0 to 50 dB
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00
Remote Command:
:CALCulate:PSTatistic:MARKer[1]|2|3|4:X <dB>
Sets the marker X position in the current X-axis units (dB).
:CALCulate:PSTatistic:MARKer[1]|2|3|4:X?
Queries the marker X position in the current x-axis units. For the response to be valid, the
marker must be on.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:X:POSition <dB>