224 Chapter 2
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
Remote Command:
[:SENSe]:SPURious[:RANGe][:LIST]:POINts[:FUNCtion]
NORMal|AVERage|POSitive|SAMPle|NEGative|QPEak|EAVerage|EPOSitive|MPositive
[:SENSe]:SPURious[:RANGe][:LIST]:DET?
Example: SPUR:POIN QPE
2.24.3.13 Detector 2
PSA Only: Allows you to set the detector to be used by the trace which will be used for
display purposes only.
The detector choices are:
Normal displays the peak of CW-like signals and maximums and minimums of
noise-like signals.
Average displays the average of the signal within the bucket. The averaging method
depends upon Avg Type selection (voltage, power or log scales).
Peak displays the maximum of the signal within the bucket.
Sample displays the instantaneous level of the signal at the center of the bucket
represented by each display point.
Negative Peak displays the minimum of the signal within the bucket.
Quasi Peak a fast-rise, slow-fall detector used in making CISPR compliant EMI
measurements.
EMI Average displays the instantaneous level of the signal at the center of the bucket,
just like the sample detector. Also changes the auto coupling of VBW, RBW and
Avg/VBW Type and the set of available RBWs. Used in making CISPR-compliant
measurements.
EMI Peak the same as the Peak detector but uses CISPR related bandwidths.
MIL Peak the same as the Peak detector but uses MIL related bandwidths.
Off turns off second trace (Detector 2)
Key Path: Meas Setup, Range Table
State Saved: Saved in instrument state.
Factory Preset: Auto, the value is determined by the radio standard and the instrument
model.
Knob Increment: 2 dB
Step Key
Increment: 2 dB
Range: 0 to 70 dB