Chapter 2 153
One-Button Measurement Functions
Meas Setup (Intermod (TOI))
One-Button Measurement Functions
2.15 Meas Setup (Intermod (TOI))
When Intermod (TOI) measurement has been selected in the Measure menu of the Spectrum
Analysis Mode, this key displays the appropriate measurement setup menu for third order
intercept (TOI).
The TOI measurement computes and displays the output intercept point (IP3), and places
markers upon the trace to indicate the measured signals and third-order products. The
span must be wide enough so that the signals and their intermod products are visible on
the display. Otherwise the TOI signal pattern cannot be identified so the measurement
routine will not function.
Key Path: Front-panel key
Dependencies/
Couplings: Menu changes depending on the Mode and Measurement selected.
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.08.00
Remote Command:
There is no equivalent remote command.

2.15.1 Avg Number

Set Avg Num ber to On to specify the number of measurement averages used when
calculating the measurement result. The average will be displayed at the end of each
sweep. Setting Avg Numb er to Off disables measurement averaging.
Key Path: Meas setup
State Saved: Saved in instrument state.
Factory Preset: 10 averages / Off
Range: 1 through 1000
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.08.00
Remote Command:
[:SENSe]:TOIN:AVERage:COUNt <integer>
[:SENSe]:TOIN:AVERage:COUNt?
[:SENSe]:TOIN:AVERage[:STATe] OFF|ON|0|1