Chapter 2 143
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
measurement was started is used for finding the fundamental.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 10 ms / Auto
Default Terminator: seconds
Range: 10 ms to maximum sweep time of the analyzer
Remote Command:
[:SENSe]:HARMonics:SWEeptime:STATe OFF|ON|0|1
[:SENSe]:HARMonics:SWEeptime:STATe?
[:SENSe]:HARMonics:SWEeptime <time>
[:SENSe]:HARMonics:SWEeptime?
Remote Command Notes: ON is equal to automatic (Auto). OFF is equal to manual
(Man).
Example: HARM:SWE:STAT OFF
HARM:SWE 100 ms
2.13.5 Range Table (On/Off)
Press Range Table to specify whether the range table is to be used or not. When Range
Tabl e is set to Off the harmonics are measured. When Range Table is set to On the first
active range is considered to be the fundamental and the subsequent active ranges are
used as the harmonics. At least two ranges must be active if Range Table is set to On.
Key Path: Meas Setup
Factory Preset: See Table 2-1 on page145.
Range: Off|On
History: Added with PSA firmware revision A.03.00
Added with ESA firmware revision A.09.00
Remote Command:
[:SENSe]:HARMonics:RTABle:STATe OFF|ON|0|1
[:SENSe]:HARMonics:RTABle:STATe?
Example: HARM:RTAB:STAT OFF