Chapter 2 111
One-Button Measurement Functions
Meas Setup (Burst Power)
One-Button Measurement Functions
points.
RMS — Selects averaging that converts trace data from dB to power units, then
averages the power trace data. This selection requires more time to perform.
Key Path: Meas Setup
Factory Preset: RMS
Range: RMS or Log
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00
Remote Command:
[:SENSe]:BPOWer:AVERage:TYPe LPOWer|POWer
[:SENSe]:BPOWer:AVERage:TYPe?
Example: BPOW:AVG:TYP LPOWer to select Log type.
BPOW:AVG:TYP?
2.5.4 Threshold Lvl
Enables you to set the level above which the mean carrier power calculation is based. The
threshold level can be described in dB (Rel) or dBm (Abs).
Key Path: Meas Setup
Factory Preset: 30 dB
Terminators: dB or dBm
Default Terminator dB
Knob Increment: 0.1 dB/dBm
Step Key
Increment: 6 dB/dBm
Range: –60 dBm to 60 dBm (in absolute mode)
–60 dB to 0 dB (in relative mode)
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00
Remote Command:
[:SENSe]:BPOWer:THReshold <number>
[:SENSe]:BPOWer:THReshold?
[:SENSe]:BPOWer:THReshold:TYPE ABSolute|RELative